Log in | Create profile | Language

Xineos-2329

Better Images, Lower Dose
for Mammography

The Xineos-2329 CD42M222 CMOS X-ray detector is specifically designed to address the very demanding needs of X-Ray mammography applications with just one detector suitable for both static FFDM (Full Field Digital Mammography) and dynamic DBT (Digital Breast Tomosynthesis) modes. The imaging performance of the detector exceeds the current industry benchmarks, while it provides several unique features.

Teledyne DALSA's advanced CMOS detector delivers unsurpassed low dose DQE at operational speeds, which allow for easy implementation of emerging modalities like DBT dual-energy imaging, etc. Proven radiation-hard design ensures stable and reliable performance over the product lifetime. A high quality CsI scintillator handles X-ray conversion, and an SDK is provided for ease of integration.

Note: This product is available only for mammography and DBT applications at this time. For other applications please contact us.

 

Feature Highlights

  • 49.5 μm square 3T PPD radiation-hard CMOS active pixel design
  • 23 x 29 cm active area
  • X-ray stack optimized for high resolution imaging in mammography, specimen analysis, etc.
  • “Dynamic saturation” for application specific settings (patent pending)
  • 14-bit image data in DBT-mode; 16-bit image data in Full Field Digital Mammography (FFDM)
  • ≤2.8mm chestwall edge distance
  • Low power consumption
  • No ghosting for typical clinical dose rates
  • GigE interface for ease of integration
  • Offset, Gain and Defect Pixel Correction (Win7 64-bit library)
  • Long life time with stable performance
  • Software to support detector factory calibrations

 

Typical Performance Values @ RQA-M2

Scintillator Medical-grade columnar CsI
Saturation Dose, RQA-M2 (per mode) 8 uGy FFDM mode / 1 uGy DBT mode
Dynamic Range (per mode) 87 dB / 73 dB
MTF @ 1lp/mm / 2lp/mm 90% / 37%
DQE(0)  70%
ADC Conversion 16-bit / 14-bit
Frame Readout Time <500 msec / <125 msec
Chest Wall Distance <2.8 mm
X-Ray Energy Range All mammography modalities
Power Consumption (Active) <20W

 